Gossen R2600 Network Card User Manual


 
GOSSEN METRAWATT GMBH 32
4.4 Equipment specifications
The equipment specifications in main group 3 of the parameter index among others include marking identification, software version and some configurations.
4.4.1 Table of equipment specifications
PI Parameter Format Unit Setting range Remarks
30h Equipment marking 8 bits 26h Read only
31h Marking identification 8 bits Bit field (4.4.2) Read only
32h Configuration
Sensor unit, continuous output
8 bits (4.4.3) CnF1, 4th digit
33h Sensor type, B marking 2x8 bits (4.4.4) CnF1, 3rd digit
35h Software version 8 bits e.g. 18h = version 1.8 Read only
36h Configuration of alarms 1 / 2 8 bits Bit field (4.4.5) CnF1, 1st digit
CnF2, 1st digit
39h Configuration of
switching outputs I / II
8 bits Bit field Value I II Read only
(DIP switch)
00h
08h
10h
18h
Transistor
Relay
Transistor
Relay
Transistor
Transistor
Relay
Relay
3Ah Continuous signal Cont 8 bits 0 = actual set point
1 = Cool reg. ratio (instead of II)
Only active, if PI = 32h
set to 8
3Fh OEM version number 8 Bit 0 = no OEM version Read only