White calibration for reflectives
e Flextight X5 scanner uses a white reference to make sure that all elements in the
scanner’s CCD react consistently to light brightnesses. e scanner has been calibrat-
ed at the factory, and the calibration data is stored in the scanner’s flash PROM.
As the scanner ages, the properties of the light tubes may change slightly. erefore,
you must redo the white calibration every six months or so and immediately aer
changing the light tubes.
Use the procedure below to make a White Calibration:
1. Find the white calibration sheet that came with the scanner. It is a square white
piece of paper (219 x 219 mm (8.6” x 8.6”)). If you have lost the white calibration
sheet, then you can use a at (non-textured), perfectly white piece of paper of the
same dimensions.
2. Your scanner came with several clear plastic layers for the reective original holder.
If you have used the reective holder before, make sure that there are no nger
prints or other marks on the plastic layer. Replace the layer with a new one if
necessary.
3. Align the top edge of the sheet with the dotted line near the top of the original
holder (see the gure below). Align each side of the target with the edges of the
original holder.
4. Select
White Calibration from the Maintenance window. The External White
Calibration window appears.
5. Click on
Calibrate. The scanner will take several scans of the white target and the
whole process will take about eight minutes.
13