HP (Hewlett-Packard) ColorFlex 13-1 Printer User Manual


 
07/02 ColorFlex 13-21
7. When all readings are complete, examine the average DE* value. The instrument passes the
repeatability test as long as the average DE* value is less than or equal to 0.04. Contact HunterLab
Customer Support as described on page 25-6 if the instrument fails the repeatability test.
Note: Performance specifications are subject to change without notice.
Regulatory Notice
A copy of the Declaration of Conformity according to ISO/IEC Guide 22 and EN 45014 follows on the