HP (Hewlett-Packard) E4407B Printer User Manual


 
Chapter 3 49
Making a Basic Measurement
Viewing a Signal
Improving Frequency Accuracy
9. While not all of the zeros following the decimal in the active function block are
significant, the numbers after the decimal in the marker annotation (upper-right
corner of screen) are significant. To increase the accuracy of the frequency reading in
the marker annotation, turn on the frequency count function.
a. Press
Mkr Fctn.
The
Marker Fctn softkeys appear.
b. Press
Marker Count.
The
Marker Count softkeys appear.
10.While not all of the zeros following the decimal in the active function block are
significant, the numbers after the decimal in the marker annotation (upper-right
corner of screen) are significant. To improve the accuracy of the frequency reading in
the marker annotation, turn on the frequency count function by pressing
Freq Count.
11.Note softkey
Marker Count On Off. If Off is underlined, press the softkey to toggle
marker count on.
As shown in Figure 3-5:
The marker annotation changes from Mkr1 to Cntr1.
The displayed resolution in the marker annotation improves.
NOTE When you use the frequency count function, if the ratio of the resolution
bandwidth to the span is less than 0.002, the following message
appears on the display: Marker Count: Widen Res BW
This indicates that the resolution bandwidth is too narrow.
12.Press
Marker , Mkr CF to move the 10 MHz peak to the center of the display.
Valid Marker Count Range
13.Move the marker down the skirt of the 10 MHz peak. Note that although the readout
in the active function changes, as long as the marker is at least 26 dB above the noise,
the counted value (upper-right corner of display) does not change (see Figure 3-6). For
an accurate count, the marker does not have to be at the exact peak.
NOTE Marker count functions properly only on CW signals or discrete peaks.
For a valid reading, the marker must be 26 dB above the noise.
14.Press
BW/Avg, Res BW, then enter a new value. This action makes the resolution
bandwidth the active function and allows you to experiment with different resolution
bandwidth (RBW) values.
15.Press
Marker, Off to turn the marker off.