13
Precautions:
• Always use locking mechanism and O-ring
• Sockets not used must be fi tted with blind plugs to ensure IP Degree
• When using the modular plug with open end, is it up to the customer to maintain
the IP degree.
• If the MJB – Gateway is used as open loop, there is a risk of not sending an alarm,
when no mains are connected. The MJB – Gateway will only send an alarm if
mains is missing, when using a Closed Loop (see functionality description.)
• When the Gateway is used on a system with battery,
the gateway functionality will follow the power
down mode of the control box, see table
• The MJB – Gateway is not to be used as safety, it is only to be used as a guided
system.
• It is important to test the specifi ed alarms is working in the correct way, before
sending the system to the end user.
• Do not use 2 MJB variants with same device ID on the OpenBus™. This will cause
confl icts and the SDT is not able to identify the different products attached.
• Before the fi nal functional test in the production, it is important that the system
is re-powered. This is to ensure that all items have been detected on the
OpenBus™.
• We recommend the end user to make a regular test procedure in order to
prevent failures and hazardous situations on the system, e.g. defective OOB
sensor, squeezed cables. The MJB is not able to detect defective 3
rd
. party products.
• LINAK only takes responsibility for LINAK products, not 3
rd
. party products.
Please pay attention to the “Patient Environment” Clause 3.79 - EN60601-1 3
rd
edition: Any volume in which contact can occur between a patient and parts of the
medical equipment or between a patient and other persons touching parts of the
equipment. Justifi ed practice here is 1.5 m. This volume must be subject to the Risk
Analysis.
It is important to inform the customer about this.
• There can be a risk of confl ict with other OpenBus™ accessories, such as HB, ACP,
etc. It is therefore recommended to make a system/bit overview.
CB Power mode Gateway
On mains Function ok
On battery
Power down
“Wake Up”
No function
Function ok