National Instruments NI 6014 Network Card User Manual


 
Appendix A Specifications
NI 6013/6014 User Manual A-2 ni.com
FIFO buffer size......................................512 samples
Data transfers..........................................DMA, interrupts,
programmed I/O
DMA modes ...........................................Scatter-gather
(Single transfer, demand transfer)
Number of DMA channels .....................1
1
Configuration memory size ....................512 words
Accuracy Information
Transfer Characteristics
Relative accuracy....................................±1.0 LSB typ, ±3.0 LSB max
DNL ........................................................±0.5 LSB typ, ±1.0 LSB max
No missing codes....................................16 bits, guaranteed
Offset error
Pregain error after calibration..........±2.0 µVmax
Pregain error before calibration.......±28.8 mV max
Postgain error after calibration ........±305 µVmax
Postgain error before calibration .....±40.2 mV max
1
The NI 6013/6014 has one DMA channel to be shared by all resources on the device.
Nominal
Range at
Full Scale
(V)
Absolute Accuracy Relative Accuracy
% of Reading Offset
Noise + Quantization
(µV)
Absolute
Accuracy at
Full Scale
(mV)
Te mp
Drift
Resolution (µV)
24 Hours 1Year (µV) Single Pt. Averaged (%/°C) SinglePoint Averaged
±10 0.0658 0.0700 1897.5 933.0 82.40 8.984 0.0010 1084.90 108.49
±5 0.0158 0.0200 959.8 466.5 41.20 2.003 0.0005 542.45 54.245
±0.5 0.0658 0.0700 115.8 56.2 5.035 0.471 0.0010 66.299 6.630
±0.05 0.0658 0.0700 31.4 31.40 3.067 0.069 0.0010 40.382 4.038
Note: Accuracies are valid for measurements after calibration. Averaged numbers assume dithering and averaging of 100 single-channel
readings. Measurement accuracies are listed for operational temperatures within ±1 °C of internal calibration temperature and ±10 °Cof
external or factory calibration temperature.