ResMed 248127 Network Router User Manual


 
17Cleaning and Maintenance
Cleaning and Maintenance
You should regularly carry out the cleaning and maintenance described in
this section.
Daily Cleaning
Mask
Clean the mask according to the instructions supplied with the mask.
Air tubing
Disconnect the air tubing from the VPAP unit (and humidifier, if used) and
hang the tubing and mask in a clean, dry place until next use.
!
CAUTION
Do not hang the air tubing in direct sunlight as the tubing may harden
over time and eventually crack.
Humidifier
If you are using a humidifier, clean it according to the instructions in the
humidifier user manual.
Weekly Cleaning
1. Remove the air tubing from the VPAP unit and the mask.
2. Wash the air tubing in warm water using mild detergent. Rinse
thoroughly, hang and allow to dry.
3. Before next use, assemble the mask and headgear according to the user
instructions.
4. Reconnect the air tubing to the air outlet and mask.
!
CAUTION
Do not use bleach, chlorine-, alcohol- or aromatic-based solutions
(including all scented oils), moisturising or antibacterial soaps to
clean the air tubing or the VPAP. These solutions may cause
hardening and reduce the life of the product.
Do not hang the air tubing in direct sunlight as the tubing may
harden over time and eventually crack.
Periodic Cleaning
1. Clean the exterior of the VPAP unit with a damp cloth and mild liquid
soap.
2. Inspect the air filter to check if it is blocked by dirt or contains holes. See
“Replacing the Air Filter” on page 18.