ResMed 248127 Network Router User Manual


 
35System Specifications
System Specifications
Dynamic pressure characteristics
IPAP: 2 cm H
2
O to 25 cm H
2
O (measured at the end of standard 2 m air
tubing)
EPAP: 2 cm H
2
O to 25 cm H
2
O (measured at the end of standard 2 m air
tubing)
CPAP: 4 to 20 cm H
2
O (measured at the end of standard 2 m air tubing)
Maximum single fault pressure: 40 cm H
2
O
Maximum flow (CPAP mode, measured at the end of standard 2 m air
tubing)
Sound pressure level: <30 dB (tested in accordance with the requirements
of ISO 17510-1:2002)
Dimensions (L x W x H): 270 mm x 230 mm x 141 mm
Weight: 2.3 kg
Air outlet: 22 mm taper, compatible with EN 1281-1:1997 Anaesthetic &
Respiratory Equipment - Conical Connectors
Pressure measurement: Internally mounted pressure transducer
Flow measurement: Internally mounted flow transducer
Power Supply: Input range 100–240V, 50–60Hz, 40VA (typical power
consumption), < 100VA (maximum power consumption)
Housing Construction: Flame retardant engineering thermoplastic
Environmental Conditions
Operating Temperature: +5°C to +40°C
Operating Humidity: 10%–95% non-condensing
Storage and Transport Temperature: -20°C to +60°C
Storage and Transport Humidity: 10–95% non-condensing
Electromagnetic Compatibility
Product complies with all applicable electromagnetic compatibility
requirements (EMC) according to IEC60601-1-2, for residential, commercial
and light industry environments. For further details, see “Guidance and
Manufacturer’s Declaration - Electromagnetic Emissions and Immunity” on
page 39.
Air Filter: Two-layered, powder-bonded, polyester non-woven fibre
Pressure
(cm H
2
O)
Flow
(L/min)
10 200
15 170
20 150