Teledyne 9950 Computer Monitor User Manual


 
Easy Calibration & Test Data Storage
An auto-zero feature for the gas sensors takes
even more of the work out of your hands. Span settings
are carried out by means of front-panel control switches,
and keypad controls remain functional during setup,
allowing the operator to select any analysis parameters
or fuels as necessary. All settings remain in memory –
even when power is turned off – and will be stored until
amended by the user or the analyzer is deliberately "cold
started" to clear the memory.
The 9950 stores up to 99 individual tests (including
a complete set of measurements and boiler ID number)
which are maintained in a non-volatile memory and can
be retrieved for printout for downloading via the RS-232
serial port.
Sample Integrity is Never in Question
Model 9950 accomplishes sample conditioning
through a fully integrated handling system capable of
delivering stack gas to the analyzer without
compromising integrity. A stainless steel probe is
connected to the monitor by a flexible hose. A high-
efficiency vane pump draws the sample through the
probe and into a coalescing filter where particulates and
condensates are removed prior to analysis, assuring you
the highest accuracy possible. Water and particulate
matter are automatically pumped from the system.
Measured Parameters
Range* Resolution Accuracy
Oxygen (02) 0-25% 0.1% +/-1%FS**
Electrochemical cell
Expected Life - 24-36 months
Carbon Monoxide 0-2000 ppm 1ppm* +/-3%FS**
Electrochemical cell
Expected Life - 24-36 months
Sulfur Dioxide (SO
2) 0-1000 ppm 1ppm* +/-3%FS**
Electrochemical cell
Expected Life - 24-36 months
Nitric Oxide (NO) 0-2000 ppm 1ppm* +/-3%FS**
Electrochemical cell
Expected Life - 24-36 months
Nitrogen Dioxide (NO
2) 0-200 ppm 1ppm* +/-3%FS**
Electrochemical cell
Expected Life - 24-36 months
Combustion Gas Temperature 0-850°C 1°C
K-type thermocouple continuous
0-1200°C 1°C
intermittent
Ambient Temperature 0-50°C 1°C
(for less than 85% relative humidity)
0-40°C
(for 85% relative humidity-printer specifications)
Calculations
Range Resolution Accuracy
Carbon Dioxide (CO2) 0-20% 0.1% +/-5%
Combustion Efficiency 0-100% 0.1% varies with application
Pollutant Emissions 0-9.999 0.001 +/-3%
lbs / million BTU
* Please contact factory regarding non-standard ranges.
** At constant temperature once thermal equilibrium is achieved.
Options Probe and hose lengths
NOTE: longer lengths may not fit within case
Selected optimal ranges for the
CO / SO
2 / NO / NO2 analysis are available
Remote display unit
TELEDYNE ANALYTICAL INSTRUMENTS
MODEL 9950 PEM
2:18 PM 1/19/98
FUEL NO.: #2 OIL
COMBU. EFF.: 70.1%
STACK TEMP.: 820˚F
O
2
: 4.8%
CO: 987 PPM
CO
2
: 11.5%
NO: 2 PPM
NO
2
: 1 PPM
NOX (NO+NO
2
): 3 PPM
SO
2
: 1 PPM
SO2 O2
NO CO
NO