A SERVICE OF

logo

To set up the source output resistance test
Performance Test only
This test verifies that the Agilent 35670A meets its source output specification for
resistance. In this test, a digital multimeter measures the 50
feedthrough
termination. The channel 1 input then measures the source output across the
feedthrough termination, then in an open circuit condition. The resistance is
calculated using the following formula:
Rs = R1((Vopen - Vload)/Vload)
Note: Use the same 50
feedthrough termination for steps 1 and 2.
1 2ch
2 2ch
Verifying Specifications Agilent 35670A
To set up the source output resistance test
3-46