A SERVICE OF

logo

To set up the cross talk test
Performance Test only
This test verifies that the Agilent 35670A meets its input specification for
channel-to-channel and channel-to-source cross talk. In this test, the
Agilent 35670A measures the amount of energy induced from the source or input
channel to another input channel. For source-to-channel crosstalk, the analyzer’s
source is set for 25.6 kHz, 9 dBVrms and the signal level at the input channels is
measured. For channel-to-channel crosstalk, the frequency synthesizer outputs a
25.6 kHz or 51.2 kHz, 9 dBVrms signal to all but one input channel and the signal
level at the unused input channel is measured.
1 2ch
2 2ch
Verifying Specifications Agilent 35670A
To set up the cross talk test
3-30