4-20 Agilent E5250A User’s Guide, Edition 8
Setting up Measurement Environment
Connecting the E5250A Output
Connecting the E5250A Output
This section describes how to connect the E5252A and E5255A output to connector
plate, your test fixture, and so on.
• “To Connect the E5252A Output”
• “To Connect the E5255A Output”
WARNING Turn off the E5250A and all instruments connected to the E5250A. And do not
turn them on until the connection described in this section is completed. If you
ignore this warning, you may be exposed to dangerous voltage.
To Connect the E5252A Output
The E5252A output connector is shown in Figure 4-8. The E5250A input (SMU
input and AUX input) appears at the E5252A output connector as shown in Figure
4-8 when input-output switching path is made. Figure 4-8 shows signals appear on
the E5252A output connectors with non-Kelvin connection. If you make Kelvin
connection, the signal of the even output connectors is Sense, not Force.
Figure 4-8 E5252A Output Connector and Output Signal
You can easily connect the E5252A output connector to the connector plate or your
test fixture by using cables shown in Table 4-4.
(a) When SMU is connected
(b) When AUX is connected
Signal (force)
GND
Signal (force)
GND
GUARD