Installation
SECTION II
INSTALLATION
2-1. INTRODUCTION
2-2. This section provides instructions for unpacking, inspection, preparation for use, ship-
ment, and storage.
2-3. UNPACKING AND INSPECTION
2-4. If the shipping carton is damaged, inspect the counter for visible damage (scratches,
dents, etc.). If the counter is damaged, notify the carrier and the nearest Hewlett-Packard
Sales and Service Office immediately (offices are listed at the back of this manual). Keep the ship-
ping carton and packing material for the carrier’s inspection.
2-5. PREPARATION FOR USE
CAUTION
Before connecting this instrument to an ac power line, be sure that
the 115—230-volt line selector switch on the rear panel is set to the
proper position and proper line fuse is installed (see below).
2-6. Power Requirements
2-7. This instrument can be operated on single phase 115 or 230 (-10%. +5%) volts ac. Power
required is approximately 100 VA maximum. To avoid instrument damage, the rear panel line
selector switch must be set to the correct position and the correct fuse (as labeled on the rear
panel) must be installed. See Section
III for rear panel features photograph. When shipped, the
switch is set to 115-volt ac operation.
2-8. Fuse Replacement and Installation
2-9. Two fuses are supplied with the instrument. The instrument is shipped with a 2.0 ampere
fuse installed for 115- volt operation. To change the instrument for 230-volt operation discon-
nect the ac power cable, set the line selector switch and install the 1.0 ampere fuse.
2-10. Power Cables
WARNING
TO PROTECT OPERATING AND SERVICING PERSONNEL, THIS
INSTRUMENT IS EQUIPPED WITH A THREE-PIN POWER RECEP-
TACLE. THE CENTER PIN OF THE RECEPTACLE CONNECTS THE
INSTRUMENT CHASSIS AND PANELS TO EARTH GROUND WHEN
USED WITH A PROPERLY WIRED THREE CONDUCTOR OUTLET
AND POWER CABLE. IMPROPERLY GROUNDED EQUIPMENT CAN
RESULT IN HAZARDOUS POTENTIALS BETWEEN EQUIPMENTS.
2-1