3M 390G Barcode Reader User Manual


 
6 7
1
Optional Visual pH Color ChangeResult
3M™Attest™ 1294Rapid Readout Biological Indicators may be further incubated for
a visual pH color change. Due to the high sensitivity of the fluorescent result, there is no
advantage to incubating the 3M™Attest™ 1294Rapid Readout Biological Indicators
beyond 4hours. This optional further incubation is not intended for routine use but may
be used for special studies. A humidified incubator will be required to avoid media dry out.
In the case of a positive control BI, a yellow color change of the growth media will appear
within 7days. Any observation of a yellow color within the vial indicates a positive result.
The pH color change result can be recorded and the BI discarded as soon as the yellow
color is observed and the result recorded. In the case of a processed BI, a media color
change from blue‑green to yellow indicates a sterilization process failure. A negative pH
color change result, i.e., media remains blue‑green, can be assessed at 7days.
INSTRUCTIONS FOR USE –Disposal
Instrument contains a lithium battery. Dispose of instrument in accordance with federal,
state, and localrequirements.
FEATURES OF THE 3M™ATTEST™ AUTO‑READER390G
AudibleAlarm
An audible alarm will sound, the LCD display backlight will flash, and a (+) symbol will
be displayed on the LCD panel directly under the incubating BI when a positive result is
detected. An alarm will also sound if a 3M™Attest™ Rapid Readout Biological Indicator
is removed from the 3M™Attest™ Auto‑reader390G well before a final result has been
detected. Data will be lost if the 3M™Attest™ Rapid Readout Biological Indicator is not
replaced in the incubation well within 10seconds.
AlarmOn/Off
The audible alarm for positive results can be turned off for a particular 3M™Attest™
Rapid Readout Biological Indicator. This is accomplished by pressing the “Buzzer Alarm
OFF” button after the alarm has sounded. This will turn off the alarm for that BI only. Any
positive 3M™Attest™ Rapid Readout Biological Indicator results subsequently detected
in other wells will result in the alarm sounding. Removing the 3M™Attest™ Rapid
Readout Biological Indicator will reactivate the alarm for that well. Pressing the “Buzzer
Alarm OFF” button will not silence the alarm for error or caution code alarm warnings. To
disable the alarm, see configurationsection.
Stackable
To save counter space, 3M™Attest™ Auto‑reader390Gs may be stacked twohigh.
InformationButton
When the Information button is pressed once the 3M™Attest™ Auto‑reader390G
displays two lines of text in the LCDdisplay.
The top row indicates (left toright):
•
3M™Attest™ Auto‑reader390G serialnumber
•
3M™Attest™ Auto‑reader390G embedded software version (VX.Y.Z)
•
Current date andtime.
The bottom row indicates (left toright):
•
The BI type configuration for wells1‑5.
•
The BI type configuration for wells6‑10.
Press the information button twice and the 3M™Attest™ Auto‑reader390G LCD panel
will display two additional lines oftext.
The top rowindicates:
•
Dynamic IP address (if unit is connected to an Ethernetjack).
The bottom row indicates (left toright):
•
Date (YYYY‑MM‑DD) and time that power was supplied to the 3M™Attest™
Auto‑reader390G.
TemperatureMonitoring
The 3M™Attest™ Auto‑reader390G contains an internal temperature self‑diagnostic
program. Should the temperature fall outside the specified range (37ºC +/‑2ºC) an error
code will be displayed. The heater block temperature may be checked by depressing the
Temperature buttononce.
EasyConfiguration
The 3M™Attest™ Auto‑reader390G may be configured using either the buttons located
on the front of the unit or using the optional 3M™Attest™ Auto‑reader390G WebApp.
Optional 3M™Attest™ Auto‑reader390G WebApp
If desired, the 3M™Attest™ Auto‑reader390G may be connected to a PC via an Ethernet
connection for additionalcapabilities:
•
Remotely view current status of BIs in the 3M™Attest™ Auto‑reader390G via PC
or mobile hand held devices. For full functionality, the use of an HTML5compliant
browser with JavaScript enabled isrecommended.
•
Remotely view and configure 3M™Attest™ Auto‑reader390Gsettings.
•
View the last 100BIrecords.
•
Print out documentation using a local or networked printer and/or save documentation
as an electronicfile.*
Note: To access the 3M™Attest™ Auto‑reader390G Web App within a secure network,
the PC or mobile device must have access to the secure network; options include
connecting via a network Ethernet cable, signing in to the facility’s wireless network, or
connecting via a virtual private network(VPN).
* Refer to your facility’s record keeping policy andprocedure.
CARE ANDCLEANING
CAUTION: Always unplug the 3M™Attest™ Auto‑reader390G and allow to
cool before cleaning.
Do not spill liquid onto or into the instrument. Do not immerse the unit in liquid.
Clean external surfaces using only the instructions provided by the manufacturer.
The instrument must be returned to the manufacturer for repair.
Clean the exterior surface of the 3M™Attest™ Auto‑reader390G by wiping it with a
damp cloth moistened with a mild detergent. Do not attempt to clean interior components.
If additional cleaning of the wells is required, call for repair orreplacement.
Note: The 3M™Attest™ Auto‑reader390G does not contain any user serviceable parts
or require any routine maintenance. Any malfunctions will be indicated by an error code.
Refer to the Troubleshooting Guide section for error codeexplanations.