Acer X1260K Projector User Manual


 
X1160 Series/X1260 Series/X1160Z Series/X1260K Refresh Condential
4-7
(the unidentied color levels on both left and
right sides should not over 8 color levels.)
- Grey level should not have abnormal color or
heavy lines.
- The PC mode functionally sure be workable
include support format with frequency and auto
detected functional will be workable
4-7 Calibration
1. YPbPr calibration
Procedure - Test equipment: video generator.
- Once main board is changed, YPbPr calibration
should be done as well.
(1) Test signal: 480P
(2) Test Pattern: SMPTE BAR
- Note
(1) Calibration pattern should be in ll screen
mode.
(2) Please refer to 4-2 Guide to Entering Service
Mode and OSD Reset.
(3) Choose and access YPbPr Calibration for
correction in service mode. Choose “menu”
to leave the service mode after all.
Check pattern - Test signal: 480P
- Test pattern: In focus II or 64 grey RGBW
* After nishing ADC adjustment, check 64 gray
RGBW pattern.
Inspection item - Color sturations
SMPTE BAR
In focus II / 64 Gray
RGBW