A&D AD-4212A/B Postal Equipment User Manual


 
26
9. FUNCTION SWITCH AND INITIALIZATION
9-1 Permit or Inhibit
The balance stores parameters that must not be changed unintentionally (e.g. Calibration data for
accurate weighing, data for adapting to the operating environment, control data for the RS-232C
interface). There are two switches for protecting the parameters. The switches can select either
“permit” or “inhibit”. The “inhibit” protects parameters against unintentional operations. There is an
additional switch for the extended functions of the AD-4212A.
Switches
Note: Only available for the AD-4212A series balance.
For details, refer to “20. EXTENDED FUNCTIONS
Operation
1 Press the ON:OFF key to turn off the display.
2 While pressing and holding the PRINT key and the SAMPLE key, press the ON:OFF key. The
balance displays p5 .
3 Press the PRINT key. Then the balance displays the function switches.
4 Set the switches using the following keys.
SAMPLE key To select the switch to change the parameter.
RE-ZERO key To change the parameter of the switch selected.
0: To inhibit changes. 1: To permit changes
PRINT key To store the new parameter and return to the weighing mode.
CAL key To cancel the operation and return to the weighing mode.