80
Display Error code Description
Unit mass error
The sample unit mass for the counting mode is too
light. Storing and using it for counting will cause a
counting error.
Add samples to reach the specified number and press
the PRINT key.
Pressing the PRINT key without adding samples will
shift the balance to the counting mode. But, to acquire
accurate weighing, be sure to add samples.
Clock battery error
The clock backup battery has been depleted.
Press any key and set the time and date. The clock and
calendar function works normally as long as the AC
adapter is connected to the balance. If this error appears
frequently, contact the local A&D dealer.
(Blinking)
Memory full
The amount of weighing data in memory has reached
the maximum capacity.
Delete data in memory to store new data. For details,
refer to “14. DATA MEMORY”.
(Illuminated)
Memory full
The amount of calibration or calibration test data in
memory has reached the maximum capacity (50 sets).
The data in memory will be deleted automatically to
store new data. For details, refer to “14. DATA
MEMORY”.
Memory type error
Type of memory set in the function table and type of
data stored are different.
For details, refer to “14. DATA MEMORY”.
EC, E00 Communications error
A protocol error occurred in communications.
Confirm the format, baud rate and parity.
EC, E01 Undefined command error
An undefined command was received.
Confirm the command.
EC, E02 Not ready
A received command can not be processed.
e.g. The balance received a Q command, but not in the
weighing mode.
e.g. The balance received a Q command while
processing a RE-ZERO command.
Adjust the delay time to transmit a command.