m
ar345 Program Manual 3 8
_
____________________________________________________________________________________
_
Tabel continued from previous page.
_______________________________________________________________________
No. Description
_______________________________________________________________________
1101 Message: "No scan modes found in nb_code ..."
1102 "Something wrong with byteorder in nb_code ..."
1105 "No suitable scanning mode found in nb_code".
Reason: File $MARTABLEDIR/mar2300.XXX and/or
$MARTABLEDIR/mar3450.XXX may be corrupted or empty.
Action: Check file sizes (73 MB and 103MB). Use command:
catmar $MARTABLEDIR/mar2300.XXX
to look at the calibration file header.
1103 Message: "Scanner serial number in nb_code differs from config"
Reason: The scanner no. in the file headers of files
$MARTABLEDIR/mar2300.XXX and/or
$MARTABLEDIR/mar3450.XXX are not identical to
$MAR_SCANNER_NO. The calibration files may not
belong to the scanner.
Action: Call.
1110 Message: "Cannot create image file"
1111 "Cannot open image file"
1112 "Error writing image array"
1115 "Error writing image header"
Reason: Insufficient disk space or invalid file permissions.
Action: Clean up disk or check data directory permissions.
1215/1241 Message: "Cannot OPEN SHUTTER at START of EXPOSURE"
1234 "SHUTTER was already OPEN at START of EXPOSURE!"
1238 "Cannot CLOSE SHUTTER at END of EXPOSURE!"
Reason: X−ray beam shutter damaged or dirty. Otherwise, the manual
shutter operation switch mounted on the base may have
been left in the "Open" state.
Action: Check shutter and manual shutter switch on the base.
Open and close the shutter manually. Is it very hard
to operate?
1239 Message: "STEPPER error during EXPOSURE!"
Reason: Problem with PHI axis movement. Possibly software related.
Action: Mail mar.spy file. Call.
_______________________________________________________________________