Exergen D-Series Scanner User Manual


 
51 Water Street, Watertown, MA 02172 USA
800-422-3006 • 617-923-9900 • Fax 617-923-9911
Consult factory or distributor for special duty models with special accuracies and
temperature ranges for OEM and volume applications.
D501
D1001
Microscanner D501&
D501RS
D1601
D1201
D-Series Handheld IR Scanners Specifications
-50 to 550F 0 - 1000F 186 - 1207F 186 - 1600F
& -45 to 287C & -18 to 540C & 86 to 653C & 86 to 871C
Kit Included Kit Included No No
+/-1% maximum of Difference between target temperature and
instrument temperature when touching, for emissivity of 0.8 to 1.0
1/4 inch (6.4 mm) approximately
None
+/-1% of Reading - maximum +/-3% maximum
Automatic Emissivity Compensation System
Bright LEDs at 10 times per second
80 msec approximately
2 to 20 microns
Temperature
Range
Relative Humidity
Measurement
Emissivity
Adjustment
Calibration
Requirement
Linearity Error
Emissivity Error
Repeatability
Resolution
Display /Update
Response Time
Field of View
Minimum Spot
Size
Spectral
Sensitivity
Analog Output*
Remote Sensor**
Instrument
Operating
Temperature
Battery Life
Dimensions
Weight
(Remote Sensor
Version)
Available on all models; standard on D1201 and 1601
Approximately 5000 Readings from on 9V Alkaline Battery
9 oz (.25 kg) approx.
7 oz (.2 kg) approx.
Optional
Model D501 D1201 D1601D1001
32 to 122
o
F (0 to 50
o
C)
+/-0.1 F & +/-0.1
o
C
o
0.1
o
F & 0.1
o
C
1:1 (53
o
approximately)
1 Mv / OF (1mV/
o
C)
F/ C Conversion
Yes Yes Yes Yes
oo
Main Case: 3 3/8 X 5 X 3/4” (8.5 X 12.5 X 2cm)
*Optional Analog Output - Specify “AO” example: Microscanner D501-AO
**Remote Sensor - Specify Model “RS” example: Microscanner D501-RS