The Instruments Feature:
• Full range resolution to 0.1°F/C
• SCAN, MAX and/or MIN modes of operation, model specific
• Fahrenheit/Celsius conversion
• A 10-second display lock
• An audible beeper to signal functional or conditional changes
• Hermetically sealed sensing system to withstand gas and plasma
sterilization, and cleaning with any hospital approved disinfectant
including bleach and alcohol.
• Pencil-like stainless steel sensor on the RS version.
• Optional disposable cover usage:
- Complete encasement with disposable sheaths for the LN
version.
- Full probe covering with disposable wraps for the LT version.
Optional Disposable Covers
The use of disposable covers with the DT-1001 LT is optional, depend-
ing on the requirements of the application. Recommended guidelines
are as follows:
Use With Disposable Cover
For absolute accuracy, minimizing the effects of emissivity and
evaporative cooling, contact with the measurement site is recom-
mended. Accordingly, when direct contact is employed, use of a
disposable probe cover is recommended.
When touching moist tissue, use of a disposable cover is required
specifically to avoid a lower temperature from the effects of evapo-
rative cooling, and to protect against the risk of cross contamina-
tion.
Use Without Disposable Cover
If the measurement site is dry, direct contact can confidently be
made without the use of a disposable cover. When the site is dry
and the precise temperature is not a prerequisite, the measure-
ment can be made without even contacting the skin. The probe
can be cleaned with any hospital approved disinfectant, even bleach
solution.
2