Fluke 43B Power Supply User Manual


 
Fluke 43B
Users Manual
42
Miscellaneous
Display
Useful screen area .............................................. 72 x 72 mm (2.83 x 2.83 in)
Resolution ............................................................................. 240 x 240 pixels
Backlight....................................................Cold Cathode Fluorescent (CCFL)
Power
External
Power Adapter....................................................................................PM8907
Input Voltage ..............................................................................10 to 21V DC
Power ............................................................................................ 5 W typical
Internal
Rechargeable Ni-Cd battery pack ........................................................ BP120
Voltage range................................................................................4 to 6 V DC
Operating Time....................................................................................4 hours
Charging Time........................................................4 hours with Fluke 43B off
12 hours with Fluke 43B on
Refresh cycle..............................................................................8 to 14 hours
Memory
Number of screen memories............................................................................20
Number of transient memories (temporary) .....................................................40
Mechanical
Height x width x depth.................................232 x 115 x 50 mm (9.1 x 4.5 x 2 in)
Weight (including battery pack).................................................... 1.1 kg (2.5 lbs)
Interface .......................................................................RS232, optically isolated
Supported Printers ............HP Deskjet
®
, Laserjet
®
, PostScript and Epson FX80.
Using HP PCL Protocol, Postscript, and Epson ESC/P Protocol.
Serial via PM9080 (optically isolated RS232 Adapter/Cable).
Parallel via PAC91 (optically isolated Print Adapter Cable, optional).
To PC.............................................................. Dump and load settings and data
Serial via PM9080 (optically isolated RS232 adapter/cable),
using SW43W (FlukeView
®
Power Quality Analyzer software).