Hach DR 2800 Network Router User Manual


 
70
Advanced Operations
6.1.1.4 Store a user program
The input of the basic data is complete. An overview of the variable
program data is displayed.
1. To enter more specifications or change existing ones, highlight
the appropriate line and press Edit.
2. Select Store to save the user program.
3. Press the PC & Printer icon to send the program data to a
printer or to a USB memory stick (connect the USB memory
stick to the USB interface first).
Note: The test data will be formatted in the preinstalled PrgData folder
as a .csv and as a .lst file.
4. To transfer the program data from one instrument to another
create a new folder on the USB memory stick named dbhlc
and copy the .lst files into this folder. (see section 6.7.2 on
page 98).
6.1.1.5 Additional user-defined parameters and functions
In addition to the previously defined basic data, additional
parameters and functions can be defined for user programs:
measuring range upper and lower limits
timer functions
chemical forms
Upper and lower limits of the measuring range
It is possible to enter a maximum (upper) and minimum (lower)
measurement value. An error message is displayed if a reading is
above the upper limit or below the lower limit.
1. Highlight the appropriate line in the overview of the program
data and press Edit.
2. Set On and press the 0.000 key to enter the measuring range
limit. Confirm the entry by pressing OK.