Reboot the NumaLink-3.0 computer. If the studies still do not come over, reboot the source and
destination systems as well. Try to determine if the data comes over to the NumaLink-3.0 system, and
also if it leaves the Numa system, but cannot be seen or processed at the destination. Check the
following:
Does this happen with only one patient, or all patients?
Does this happen with only one study type, or all study types?
If the problem only happens with one patient, is there anything different in the acquisition (different
parameters than normal)?
If the problem is with one study type, has the transfer of that type worked before? Do other patients
who have had the same study come across?
This information will assist the Numa Technical Staff in the troubleshooting process
Studies that I sent to my ADAC Pegasys computer do not show up in the patient list, even though they
were successfully transferred.
Your Pegasys system may have the incorrect date range set, and the studies you sent may be out of range.
Adjust the date range field in the patient list on the Pegasys to reflect the acquisition date of the study you
sent. You may also need to specify a date in the future if the date on the ADAC system is incorrect.