SCANNER OPERATION
INDICATOR DESCRIPTIONS
Diagnostic Indicator Display
ERROR
CODE
DESCRIPTION
13
SCALE ERROR
The scanner does not communicate with the scale. Make sure there is nothing plugged into the ‘Scale RS232 to
Host’ port. If there is disconnect the cable, it may be causing the communication error. If there is still no
communication between the scanner and scale return the unit to authorized service center for repair.
14
SCALE RETURN TO ZERO ERROR – The scale did not return to zero between scale weight requests if the zero
checking function is enabled. The scale will need to be re-zeroed.
21
LASER #1 (VERTICAL) ERROR – The laser in the vertical scanning subsystem denotes a failure. The scanner will
try three times to correct the laser operation. If the laser error persists, the vertical scanning subsystem will be shut
down and this error code will remain on the Diagnostic Indicators.
22
LASER #2 (RIGHT HORIZONTAL) ERROR – The right laser in the horizontal scanning subsystem denotes a
failure. The scanner will try three times to correct the laser operation. If the laser error persists, and the left
horizontal laser (#3) is also in error, the horizontal scanning subsystem will be shut down and this error code will
remain on the Diagnostic Indicators.
23
LASER #3 (LEFT HORIZONTAL) ERROR – The left laser in the horizontal scanning subsystem denotes a failure.
The scanner will try three times to correct the laser operation. If the laser error persists, and the right horizontal
laser (#2) is also in error, the horizontal scanning subsystem will be shut down and this error code will remain on the
Diagnostic Indicators.
24
Laser #1 (Vertical) High Current Warning – The laser in the vertical scanning subsystem is drawing excessive
current. This could be a warning for a future total failure. Have the unit checked at an authorized service center
when convenient.
25
Laser #2 (Right Horizontal) High Current Warning – The laser in the vertical scanning subsystem is drawing
excessive current. This could be a warning for a future total failure. Have the unit checked at an authorized service
center when convenient.
26
Laser #3 (Left Horizontal) High current Warning – The laser in the vertical scanning subsystem is drawing excessive
current. This could be a warning for a future total failure. Have the unit checked at an authorized service center
when convenient.