Metrologic Instruments MS2320 Scanner User Manual


 
41
SCALE OPERATION
CALIBRATION
Scale Calibration Procedure (lbs. or kg) with Remote Display*
5. Enter calibration mode. The remote display will flash all of the characters available (see illustration below). Press the
right arrow button twice ()() while the display flashes all characters. The remote display should then read CAL 1.
Figure 44. Enter Calibration Mode
6. Reinstall the platter onto the scanner/scale. Check the platter to ensure that it is seated properly and nothing is interfering
with its freedom to move.
Figure 45. Install Platter
7. Make sure there is no load on the scale platter.
8. Calibrate a zero load. Press the right arrow () button once when the remote display reads CAL 1 and there is no load
on the scale platter.
Figure 46.
* These procedures are for MS2320 StratosH Scanner/Avery Scale models only.