42
SCALE OPERATION
CALIBRATION
Scale Calibration Procedure (lbs. or kg) with Remote Display
9. Place a half load (15.00 lbs. or 7.500 kg) on the center of the scale platter then press the right arrow () button.
Figure 47.
10. Add an additional half load (15.00 lbs. or 7.500 kg) to the existing half to simulate a full load, center entire load then press
the right arrow () button.
Figure 48.
11. Remove half of the load (15.00 lbs. or 7.500 kg), center the remaining load then press the right arrow () button.
Figure 49.
12. Remove the remaining half load from the scale then press clear (C). The message done will flash briefly on the display.
Figure 50.
13. Exit calibration mode. Press and hold the test button for at least 3 seconds then release. With no load on the scale the
display should read 0.00 lb. or 0.000 kg.
Figure 51.