National Instruments NI 9236 Network Card User Manual


 
© National Instruments Corp. 33 NI 9235/9236
Shock and Vibration
To meet these specifications, you must panel mount the system and
use the NI 9965 backshell to protect the connections.
Operating vibration
Random (IEC 60068-2-64).........5 g
rms
, 10 to 500 Hz
Sinusoidal (IEC 60068-2-6) .......5 g, 10 to 500 Hz
Operating shock
(IEC 60068-2-27)..............................30 g, 11 ms half sine,
50 g, 3 ms half sine,
18 shocks at 6 orientations
Electromagnetic Compatibility
This product is designed to meet the requirements of the following
standards of EMC for electrical equipment for measurement,
control, and laboratory use:
EN 61326 EMC requirements; Industrial Immunity
EN 55011 Emissions; Group 1, Class A
CE, C-Tick, ICES, and FCC Part 15 Emissions; Class A