Toshiba TIC-LF494B Network Card User Manual


 
TIC-LF494B
9
Terminal board
Connected detector
IV wire
5.5 mm
2
or more
Power switch
(External double-pole power switch )
Thick
w
alled steel conduit
Signal cable
Power cable
(CVV)
Digital input cable
(CVV-S)
Current output (4~20mAdc)
or PROFIBUS
Grounded with 100 or less
ground resistance
(2-wire shielded hard-rubber
sheathed cable)
(3-wire shielded hard-rubber cable)
Digital output 1
Signal common for DI and DO
Digital output 2
Digital input
(20~30Vdc)
Power supply
[Instrument panel : ordered separately]
Figure 10. Separate type LF490/LF612 and LF494/LF612F flowmeters wiring Diagram
Table 1. LF610, LF610F, LF612 and LF612F Converters Signal Table
Symbol Description Cable
L1 ()
L2 ()
Power supply Power cable (CVV)
GND Ground (for arrester)
FG Frame ground
DI Digital Input (2030Vdc)
DO1 Digital Output 1
DO2 Digital Output 2
COM Signal Common for DI, DO1, DO2
I/O cable (CVV-S)
Current Output (420mAdc)
or PROFIBUS
Shielded cable for
PROFIBUS-PA
X
Y Excitation Output
E
Excitation cable
(for LF612, LF612F only)
A
B Signal Input
G
Signal cable
(for LF612, LF612F only)
Separate type LF490/LF612 and LF494/LF612F flowmeters
Grounded with 100 or
less ground resistance
Excitation cable
Grounded with 100 or
less ground resistance