4. Operation
2. Use or to set the maximum allowed set tempe-
rature. The factory pre-set value is 90.0 °C.
3. Confirm with before the display returns to the actu-
al temperature value.
Note: The new value is not saved until has been pres-
sed.
1. Press to select the S mode.
2. Use or to set the working temperature.
3. Confirm with before the display automatically
returns to the actual temperature value.
Note: The new set value is not saved until has been
pressed.
The display continuously indicates the actual temperature in
the bath at the position of the sensor. This temperature does
not correspond directly to the temperature in the externally
connected equipment. Under most circumstances this tempe-
rature difference is very small. To adjust the display value to
reflect the temperature in the external system, proceed as fol-
lows:
1. Measure the temperature in the external system.
2. Calculate the temperature difference between this value
and the value displayed.
T
diff
= T
disp
-T
meas
T
diff
is the correction value “c” that
has to be entered.
3. Press to select the C mode
4. Use or to set the correction value with its + or
- sign. The range is -25.0 to +2.50 °C. The factory pre-set
value is 0.0.
11
4.4 Tempera-
ture setting
4.5 Tempera-
ture
correction