Xantrex Technology XG 850 Power Supply User Manual


 
Analog Programming (APG) and Isolated Analog Programming (ISOL)
4-26 M370078-01
Voltage-Controlled Current ISOL Setup
Activating ISOL Programming Current Mode
1. Turn the 9-position mode control to the CAP position or press the
rotary adjust/Enter control if the control knob is already at the CAP
position.
CAPr is displayed on the output voltage display.
2. Turn the rotary Adjust/Enter control to select U IS (voltage source
from isolated connector).
3. Press the rotary adjust/Enter control to commit the selected mode.
LE C is displayed on the output voltage display.
4. Select the input source level using the rotary adjust/Enter control.
5. Once the desired level has been selected, press the rotary adjust/
Enter control to commit the change.
The display will return to the normal display.
Deactivating ISOL Current Mode
To deactivate the ISOL current mode:
1. Turn the 9-position mode control to the VAP position or press the
rotary Adjust/Enter control if the control knob is already at the VAP
position.
2. CAPr is displayed on the output voltage display.
3. Turn the rotary Adjust/Enter control to select nonE.
4. Press the rotary Adjust/Enter control to commit the change.
The display will return to the normal display.