Xantrex Technology XPD 60-9 Power Supply User Manual


 
Section 1. Features and Specifications
Additional Specifications
Release A 19
Additional Specifications
Input Conditions
1. Front output current limited to 30 A maximum.
2. For input voltage variation over the AC input voltage range, with constant rated load.
3. For 0 to 100% load variation, with constant nominal line voltage (rear output only).
4. Current mode noise is measured from 10% to 100% of rated output voltage, full current, resistive load.
5. Maximum drift over 60 minutes with constant line, load, and temperature, after power on.
6. Maximum drift over 8 hours with constant line, load, and temperature, after 60-minute warm-up.
7. Change in output per ° C change in ambient temperature, with constant line and load.
8. Typical efficiency is measured at 120 V and full output power.
Rise Time (No load, full load) 50 ms
Fall Time (Full load) 50 ms
Fall Time (No load) 400 ms
Voltage Mode Transient Response 1 ms
Time delay from power on until output stable 3 s maximum
Rated AC Input Voltage 100/120/200/230/240 Vac, universal input
Maximum AC Input Power 700 W
Operational AC Input Voltage Range 85 to 264 Vac; power factor corrected
Maximum Input Current 7 A maximum at 100 Vac, 6 A maximum at
120 Vac, 3 A maximum at 220 Vac
Input Frequency Range 47 to 63 Hz
Power Factor 0.98 minimum for full load at nominal
voltage
Input Harmonic Distortion Current harmonics meet IEC 1000-3-2