Xerox 8825 Copier User Manual


 
XEROX WIDE FORMAT COPY SYSTEM Main Menu 135
Select the CALIBRATION MOTOR option to set the calibration zone
motor.
To run the CALIBRATION MOTOR diagnostic:
1. Press the
Menu key. The MAIN MENU screen is displayed.
2. Press the
Next key until DIAGNOSTICS is highlighted.
3. Press the
Enter key to select DIAGNOSTICS. The
DIAGNOSTICS menu screen is displayed.
4. Press the
Next key until SCANNER DIAGNOSTICS is displayed.
5. Press the
Enter key to select SCANNER DIAGNOSTICS. The
SCANNER DIAGNOSTICS menu screen is displayed.
6. Press the
Next key until COMPONENT TESTS is highlighted.
7. Press the
Enter key to select COMPONENT TESTS. The
COMPONENTS TESTS menu screen is displayed.
8. Press the
Next key until CALIBRATION MOTOR is highlighted.
9. Press the
Enter key to select CALIBRATION MOTOR. The
CALIBRATION MOTOR screen is displayed.
CALIBRATION
MOTOR: OFF
FUNCTION:
ROTATE
Calibration Screen
10. Press the
Previous or Next key to select FUNCTION.
11. Press the
Enter key to select FUNCTION.
12. Press the
Previous or Next key to highlight the desired function.
The choices are GO TO WIDTH, GO TO CALIBRATE, GO TO
HOME, FIND HOME, and ROTATE.
13. Press the
Enter key to confirm your selection.
14. Press the
Previous or Next key to highlight MOTOR.
15. Press the
Enter key to confirm your selection.
16. Press the
Previous or Next key to highlight ON. The choices
are ON and OFF. The requested operation will execute.
If a GO TO or FIND function has been selected, the MOTOR field will be
automatically set to OFF when the MOTOR field is once again selected.
If the ROTATE function has been selected, the motor will run
continuously until you set the MOTOR field to OFF.
NOTE: If the Scan System encounters a failure accessing the calibration
motor function, the fourth line of the display will read "FAILURE CODE:"
followed by a two-digit hexadecimal number. The fourth line may also
CALIBRATION
MOTOR