Agilent Technologies 3138A-00101 and above Computer Hardware User Manual


 
General Information
11
Specifications and Supplemental Characteristics
Table 1-1 lists the specifications and supplemental characteristics of the MPS mainframe. Module specifications are
included in the appropriate module User’s Guide. Specifications are warranted in the temperature range of 0 to 55 degrees
C. Supplemental characteristics are not warranted, but are descriptions of minimum, maximum, or average performance
that has been determined by design or type testing.
Table 1-1. Specifications and Supplemental Characteristics
PERFORMANCE SPECIFICATIONS
AC Input Ratings
Standard:
87- 132 Vac
174- 250 Vac
Frequency tolerance:
47 - 63 Hz
SUPPLEMENTAL CHARACTERISTICS
Maximum AC Line Current Ratings (with 8 power supply modules operating at maximum output)
100 Vac:
29 A rms
120 Vac:
25 A rms
200 Vac:
16 A rms
220/230 Vac:
16 A rms
240 Vac:
15 A rms
Maximum power input:
3200 VA; 1800 W maximum; 1600 W typical
GPIB Interface Capabilities
Languages:
Standard Commands for Programmable Instruments (see module
Programming Guide)
IEEE 488.2 Capabilities:
SH1, AH1, T6, L4, SR1, RL1, PPO, DC1, DT1, El, and CO
Serial Link Capabilities (Multiple mainframes sharing one GPIB address)
Number of mainframes:
2
Number of modules:
16
Serial-linked power supplies:
total number of modules and serial-linked power supplies (Agilent
667xA/665xA/664xA/603xA) cannot exceed 16
Chain length:
30 m (100 ft) maximum
TrigIn/TrigOut Characteristics
Maximum Ratings:
± 16.5 V between center conductor and chassis
Trig Out (HC TTL output):
V
OL
@ 4 ma = 1.34 V maximum
V
OL
@ 2 ma = 0.8 V maximum
V
OH
4 ma = 2.66 V minimum
V
OH
2 ma = 3.18 V minimum
Trig In (10k pull-up):
I
I
L
@ 4 V = 700 µA
I
IH
@ 5.25 V = 200 µA
V
I
L
= 0.8 V maximum
V
IH
= 2.0 V minimum