Agilent Technologies 60501B Laminator User Manual


 
12
Transient Operation and Slew Circuit Test
This test verifies transient and slew circuit operation. The slew circuits cannot be calibrated. If slew rise time and/or fall
time are not within specifications or the slew circuits are inoperative, perform either the "Transient Generator
Troubleshooting", or the "Slew Circuit Troubleshooting" in Chapter 3 of the mainframe Service Manual.
a. Use the test setup of Figure 2-1 except connect an oscilloscope across the 0.10 current monitor resistor in place of the
DMM. Set power supply for:
60501B: 5V and 5A.
60502B: 10V and 10A.
b. Recall the factory default values by pressing
.
c. Select the low current range by pressing
60501B:
.
60502B:
.
d. Set up transient operation by pressing
, then
60501B:
.
60502B:
.
e. Set the slew rate to .05A/µs (50A/ms) by pressing
(blue shift key), then
, then
.
f. Adjust the oscilloscope for a single rise or fall time display. Use delayed sweep. The rise time when measures from
10% to 90% or the fall time when measured from 90% to 10% should be between:
60501B: 30 and 50µs.
60502B: 75 and 125µs.
Note that the Electronic Load's Tran annunciator is on.
60501B: 10µs/DIV
60502B: 20µs/DIV
SLEW 0.05
g. Set the slew rate to .0025A/µs (2.5A/ms) by pressing (blue shift key), then
, then
.