Agilent Technologies 6633B Power Supply User Manual


 
Language Dictionary - 4
57
OUTPut:RELay
Agilent 66332A, 6632B, 6633B, 6634B, 6611C, 6612C, 6613C, 6614C Only
This command is only valid for units with Relay Option 760, otherwise an error will occur. Programming
ON closes the output relay contacts; programming OFF opens them. The relay is controlled independently
of the output state. If the dc source is supplying power to a load, that power will appear at the relay
contacts during switching.
Command Syntax
OUTPut:RELay[:STATe]<bool>
Parameters
0 | 1 | OFF ON
*RST Value
0
Examples
OUTP:REL 1OUTP:REL OFF
Query Syntax
OUTPput:RELay?
Returned Parameters
0 | 1
Related Commands
OUTP *RCL *SAV
OUTPut:RELay:POLarity
Agilent 66332A, 6632B, 6633B, 6634B, 6611C, 6612C, 6613C, 6614C Only
This command is only valid for units with Relay Option 760, otherwise an error will occur. Programming
NORMal causes the output relay polarity to be the same as the dc source output. Programming REVerse
causes the relay output polarity to be opposite to that of the dc source output. If OUTPut = ON when either
command is sent, the output voltage is set to 0 during the time that the relays are changing polarity.
Command Syntax
OUTPut:RELay:POLarity<CRD>
Parameters
NORMal | REVerse
*RST Value
NORM
Examples
OUTP:REL:POL NORM
Query Syntax
OUTPput:RELay:POLarity?
Returned Parameters
NORM | REV
Related Commands
OUTP *RCL *SAV
OUTPut:RI:MODE
This command selects the mode of operation of the Remote Inhibit protection. The RI mode is stored in
non-volatile memory. The following modes can be selected:
LATChing
causes a TTL low signal on the INH input to disable the output. The only way
to clear the latch is by sending an OUTPut:PROTection:CLEAR command
while the INH input is false.
LIVE
allows the INH input to disable the output in a non-latching manner. In other
words, the output follows the state of the INH input. When INH is low true, the
output is disabled. When INH is high the output is not affected.
OFF
the INH input is disabled.
Command Syntax
OUTPut:RI:MODE <mode>
Parameters
LATChing | LIVE | OFF
Examples
OUTP:RI:MODE LIVE
Query Syntax
OUTPut:RI:MODE?
Returned Parameters
<CRD>
Related Commands
OUTP:PROT:CLE