Agilent Technologies 6633B Power Supply User Manual


 
4 - Language Dictionary
70
SYSTem:LOCal
For RS-232 Operation Only
This command places the dc source in local mode during RS-232 operation. The front panel keys are
functional.
Command Syntax
SYSTem:LOCal
Parameters
None
Example
SYST:LOC
Related Commands
SYST:REM SYST:RWL
SYSTem:REMote
For RS-232 Operation Only
This command places the dc source in remote mode during RS-232 operation. This disables all front
panel keys except the Local key. Pressing the Local key while in the remote state returns the front panel to
the local state.
Command Syntax
SYSTem:REMote
Parameters
None
Example
SYST:REM
Related Commands
SYST:LOC SYST:RWL
SYSTem:RWLock
For RS-232 Operation Only
This command places the dc source in remote mode during RS-232 operation. All front panel keys
including the Local key are disabled. Use SYSTem:LOCal to return the front panel to the local state.
Command Syntax
SYSTem:RWLock
Parameters
None
Example
SYST:RWL
Related Commands
SYST:REM SYST:LOC
*IDN?
This query requests the dc source to identify itself. It returns a string composed of four fields separated by
commas.
Query Syntax
*IDN?
Returned Parameters
<AARD> Field Information
Agilent Technologies Manufacturer
xxxxxA model number followed by a letter suffix
nnnnA-nnnnn 10-character serial number or 0
<A>.xx.xx Revision levels of firmware.
Example
AGILENT,66312A,0,A.00.01