Agilent Technologies 6633B Power Supply User Manual


 
4 - Language Dictionary
60
[SOURce:]VOLTage:ALC:BANDwidth?
[SOURce:]VOLTage:ALC:BWIDth?
Agilent 66332A, 6631B, 6632B, 6633B and 6634B Only
These queries return the setting of the output mode switch. The output mode switch is used to connect or
disconnect the the output capacitor located inside the unit. The returned value is 15,000 if the switch is set
to Normal and 60,000 if the switch is set to Fast.
Query Syntax
[SOURce]:VOLTage:ALC:BANDwidth?
[SOURce]:VOLTage:ALC:BWIDth?
Examples
VOLT:ALC:BAND? VOLTAGE:ALC:BWIDth?
Returned Parameters
<NR3>
[SOURce:]VOLTage:TRIGger
This command sets the pending triggered voltage level of the dc source. The pending triggered level is a
stored voltage value that is transferred to the output terminals when a trigger occurs. In order for a trigger
to occur, the trigger subsystem must be initiated (see the INITiate command in the trigger subsystem).
Command Syntax
[SOURce][:VOLTage[:LEVel]:TRIGgered[:AMPLitude]<NRf+>
Parameters
see Table 4-3
Default Suffix
V (volts)
*RST Value
0
Examples
VOLT:TRIG 20 VOLTAGE:LEVEL:TRIGGERED 20
Query Syntax
[SOURce]:VOLTage[LEVel]:TRIGgered[:AMPLitude]?
Returned Parameters
<NR3>
Related Commands
VOLT *RST
[SOURce:]VOLTage:PROTection
This command sets the overvoltage protection (OVP) level of the dc source. If the output voltage exceeds
the OVP level, then the dc source output is disabled and the Questionable Condition status register OV bit
is set (see chapter 3 under Programming the Status Registers). An overvoltage condition can be cleared
with the OUTP:PROT:CLE command after the condition that caused the OVP trip is removed. The OVP
always trips with zero delay and is unaffected by the OUTP:PROT:DEL command.
Command Syntax
[SOURce]:VOLTage:PROTection[:LEVel]<NRf+>
Parameters
see Table 4-3
Default Suffix
V (volts)
*RST Value
MAX
Examples
VOLT:PROT 21.5 VOLT:PROT:LEV MAX
Query Syntax
[SOURce]:VOLTage:PROTection[:LEVel]?
Returned Parameters
<NR3>
Related Commands
OUTP:PROT:CLE OUTP:PROT:DEL