[:ROUTe]:SYMBol:PORT
[:ROUTe]:SYMBol:PORT
In the GPIB local mode, the B2200 uses 01 to 14 (two digits numbers) to specify the
input ports 1 to 14 respectively. This command specifies a symbol string and lets the
B2200 use the specified string instead of the two digits number. The query returns
the symbol string.
Reboot and *RST clear the symbol strings. After that, the initial symbol string (two
digits numbers 01 to 14) will be set.
:SYSTem:CPON does not clear the symbol strings.
Syntax [:ROUTe]:SYMBol:PORT port_number,'symbol_string'
[:ROUTe]:SYMBol:PORT? port_number
Query response symbol_string <newline><^END>
Example OUTPUT @Agb2200;":ROUT:SYMB:PORT 1,'SMU1'"
OUTPUT @Agb2200;":ROUT:SYMB:PORT? 1"
ENTER @Agb2200;A$
This example assigns the string SMU1 to the input port 1. After that, SMU1 is used
instead of 01. In this example, A$ will be SMU1.
Parameter Explanation
port_number Input port number: 1 to 14.
symbol_string Symbol string set to the specified input port. Maximum 6
characters.
Numeric characters, alphabetic characters; upper case and
lower case, space, and the following symbols are available.
! " # $ % & ’ ( ) * + , - . / > ? : ~ ; < = > ? @ [ \ ] ‘ { | } ^ _