Agilent Technologies B2200A Computer Accessories User Manual


 
Agilent B2200 User’s Guide, Edition 2 3-9
Front Panel Operation
Operation
To Use Couple Mode
1. Press the Couple Mode key to set the couple mode ON. Pressing the key again
sets the mode OFF.
2. Perform the input couple port detection.
a. Press the
Menu key. The setup menu will be displayed on the LCD.
b. Move the cursor to SCAN, then press the
Enter key. The LCD displays NO.
c. Press the arrow key to display YES, then press the
Enter key.
d. Press the
Exit key.
3. Change the couple ports (multiple ports can be set).
a. When the couple mode is ON, press the
Port Function key. The blink-cursor
appears on the first line of the LCD.
b. Press the right/left arrow key to move the cursor on the input port number
(01 to 14) that will be set to the input couple port.
c. Press the up/down key to display C.
d. Repeat b and c for the all couple ports, then press the
Enter key.
To Display Firmware Revision
1. Press the Menu key. The setup menu will be displayed on the LCD.
2. Move the cursor to CONFIG, then press the
Enter key.
3. Move the cursor to REVISION, then press the
Enter key. The firmware revision
is displayed.
To Display Module Information
1. Press the Menu key. The setup menu will be displayed on the LCD.
2. Move the cursor to CONFIG, then press the
Enter key.
3. Move the cursor to UNIT, then press the
Enter key. The module information is
displayed.
4. Press the arrow keys to display another module information item.