Agilent Technologies FS2101 Webcam User Manual


 
RESRVD 0100
RESRVD 0101
MEM_RD 0110
MEM_WR 0111
RESRVD 1000
RESRVD 1001
CON_RD 1010
CON_WR 1011
MEMRDM 1100
DAC_CY 1101
MEMRDL 1110
MEMWRI 1111
I/O_XACTIONS 001X
MEM_XACTIONS 011X
CONFIG_XACTIONS 101X
Theory of Operation
The PCI Active Analysis Probe is a universal PCI short card that
attaches to Agilent logic analyzers. The Analysis Probe has five
major parts:
1. The input buffers
2. The latching buffers
3. The interface to the logic analyzer
4. The clocking and cycle bit generation logic
5. The extender card connector
The Input Buffers
The input buffers present a single electrical load on the PCI bus
and are made up of IDT162260 tri-port buffers. The PCI clock is
buffered by a Motorola 807 high speed clock buffer.
When the Analysis Probe is in Timing mode output Port 2 of the
tri-port buffers goes directly to the logic analyzer input
terminators. When in State mode output Port 2 is tri-stated.
Output port 1 is always enabled and goes to the latching buffers.