Datalogic Scanning DX30 Barcode Reader User Manual


 
REFERENCES
5
5.10 BATTERY SELECTION
Battery selection is required only when the Gryphon™ Mx30 reader has an Alkaline
battery and you want to use OM-Gryphon™ or C-Gryphon™ either for serial
configuration, software upgrades or to hold Gryphon™ Mx30. Since this type of
battery must not be charged it is necessary to disable the OM-Gryphon™ or
C-Gryphon™ charge function by following the procedure:
With the Gryphon™ Mx30 read the following code:
1.
Alkaline
Ì$+RN4$-]Î
The green LED on the Gryphon™ Mx30 will blink, signaling the reader has
accepted the command.
2.
Place the reader onto the charger within 10 seconds. The green LED turns
off and a short beep is emitted
To enable the charge function repeat step 1 and 2 substituting the "Alkaline" code
with the following one:
NiMh
Ì$+RN6$-gÎ
CAUTION
Attempts to charge Alkaline batteries could cause leakage of liquid,
generation of heat or, in extreme cases, explosion. If using Alkaline
batteries, carefully follow the procedure above to avoid damage.