Hasselblad Flextight 646 Power Supply User Manual


 
Resolutions
True Resolutions
The table below shows the max. resolutions (in ppi) available with each
of the original holders when scanning in True Resolutions mode.
Calibrating the Scanner
The scanner requires two types of calibration:
Focus calibration adjusts the positioning of the scanner's optics for
each zoom level. If your scans are coming out unfocused or at slightly
the wrong size, then you may need to make a new focus calibration to
adjust the zoom mechanism. The focus calibration is stored in the
scanner's flash PROM, so it is preserved when you move the scanner
to a new computer. See "Focus Calibration" on page 21 for more
information.
White calibration is only required for reflective scans (it is done
automatically for transparencies). During calibration, the scanner
scans a white target to establish the white point along the full length
of the scanner's CCD. The white calibration is stored in the scanner's
flash PROM, so it is preserved when you move the scanner to a new
computer. See "White Calibration for Reflectives" on page 22 for more
information.
Both of the calibration procedures are simple and are initiated from the
FlexColor program.
User's Guide - Flextight 646
Hardware Reference - Flextight 646
20
Resolutions
English
Original
Max. resolution (PPI)
at full scanning width.
35 mm Portrait
6300
35 mm Landscape 5000
45 x 60 mm Portrait * 4000
60 x 60 - 60 x 70 mm 3200
4" x 5" Portrait 2040
13 x 18 Portrait * 1600
A4 Portrait * 960
*) Not included in the standard delivery
Table 1: True Resolutions