Hasselblad X5 Scanner User Manual


 
Warnings and restrictions 
• Read all of the included documentation before attempting to install and use the
scanner.
• Do not touch the originals and/or the original holder while scanning.
• Do not start scanning or previewing until an original holder with an original has
been mounted.
• e original holder is only to be mounted or removed when the drum is in the load
position.
• When you turn on the scanner, the drum will roll to the load position, if it is not
there already. Do not touch the scanner while the drum is rotating.
• Do not place your fingers or any other object into the scanner while it is connected
to power.
• Before servicing or opening the scanner, the power supply must be disconnected
from the mains (unplugged). It is not sufficient simply to press the on/off button.
• Install the scanner in a location where children can not get to it. It contains small
openings and moving parts that can cause injury.
System requirements 
Below are the basic hardware requirements for the PC or Macintosh system to which
the scanner is to be connected. For information about the processor, operating
system, RAM and harddisk requirements please refer to the “Soware Reference”
manual, that comes with the FlexColor soware.
• Screen resolution of minimum 800 x 600 pixels with true colors (24-bits).
• Mouse or other pointing device.
• FireWire interface.
  Preventive maintenance 
All Flextight scanners should be serviced every 25,000 scans or every 12 months,
whichever comes first. If a Feeder is mounted, the recommended number of scans is
15,000 provided that the feeder is mounted for all scans. Please note that previews are
also counted as scans.
Newer Flextight scanners have a counter installed so that the number of scans can be
monitored from FlexColor. Please refer to the FlexColor manual to learn more about
monitoring the number of scans.
Failures or faults originating from lack of service and/or daily maintenance are not
covered by the factory warranty.