HP (Hewlett-Packard) 8753ET Network Card User Manual


 
24
Accessories
Calibration kits
Vector accuracy enhancement procedures require that
the systematic errors of the measurement system be
characterized by measuring known devices (standards)
on the system over the frequency range of interest.
Return loss specifications or typical values are provid-
ed where available for the terminations and adapters.
HP 85031B 7-mm calibration kit
Contains precision 7-mm standards used to calibrate
the HP 8753ES for measurement of devices with preci-
sion 7-mm connectors.
Includes HP part number
7 mm short/open circuit 85031-60001
7 mm 50-ohm terminations
(two each) 00909-60008
Specifications for terminations
DC to 5 GHz RL 52 dB
5 to 6 GHz RL 46 dB
HP 85032B 50-ohm type-N calibration kit
Contains precision 50-ohm type-N standards used to
calibrate the HP 8753ET/ES and 50-ohm test sets for
measurement of devices with 50-ohm type-N connec-
tors. Precision phase-matched 7-mm to type-N
adapters are included for accurate measurements of
non-insertable devices. Option 001 deletes the 7-mm
to type-N adapters, which are not needed with the
HP 8753ET.
Includes HP part number
N-male 50-ohm termination 00909-60009
N-female 50-ohm termination 00909-60010
N-male short circuit 85032-60008
N-female short circuit 85032-60009
N-female open circuit 85032-60012
N-male open circuit 85032-60007
7-mm to N-male adapter
(two each) 85054-60009
7-mm to N-female adapter
(two each) 85054-60001
Specifications for terminations:
DC to 3 GHz RL 49 dB
2 to 3 GHz RL 46 dB
3 to 6 GHz RL 40 dB
Typical adapter characteristics:
DC to 6 GHz RL 30 dB
HP 85033D 3.5-mm calibration kit
Contains a set of precision 3.5-mm standards to cali-
brate the HP 8753ES and 50-ohm test sets for the
measurement of devices with precision 3.5-mm and
SMA connectors. Precision phase-matched 7-mm to
3.5-mm adapters are included for accurate measure-
ments of non-insertable devices. Option 001 deletes
the 7-mm to type-N adapters.
Includes HP part number
3.5-mm-male 50-ohm termination 85033-60009
3.5-mm-female 50-ohm termination 85033-60010
3.5-mm-female short 85033-60014
3.5-mm-male short 85033-60013
3.5-mm-female open 85033-60012
3.5-mm-male open 85033-60011
7-mm to 3.5-mm female
adapter (two) 1250-1747
7-mm to 3.5-mm male
adapter (two) 1250-1746
Specifications for terminations:
DC to 1.3 GHz RL 46 dB
1.3 to 3 GHz RL 44 dB
3 to 6 GHz RL 38 dB
Typical adapter characteristics:
DC to 6 GHz RL 34 dB
HP 85036B 75-ohm type-N calibration kit
Contains a set of precision 75-ohm type-N standards to
calibrate the HP 8753ES with a 75-ohm test set for
measurement of devices with 75-ohm type-N connec-
tors. Precision phased-matched adapters are included
for accurate measurements of non-insertable devices.
Includes HP part number
N-male 75-ohm termination 00909-60019
N-female 75-ohm termination 00909-60020
N-female 75-ohm short 85036-60011
N-male 75-ohm short 85036-60012
N-female open 85032-20001
N-male open 85032-60007
N-male to N-male 75-ohm adapter 85036-60013
N-female to N-female 75-ohm adapter 85036-60014
N-male to N-female 75-ohm adapter 85036-60015
Specifications for terminations:
DC to 2 GHz RL 46 dB
2 to 3 GHz RL 40 dB