HP (Hewlett-Packard) 8753ET Network Card User Manual


 
7
System performance summaries
HP 8753ES (75-ohm systems)
Type-N test ports
Configuration
Network analyzer HP 8753ES Option 075
Calibration kit HP 85036B
Test-port cables HP 11857B
System dynamic range
30 kHz to 50 kHz 68 dB
1
50 kHz to 300 kHz 90 dB
1
300 kHz to 16 MHz 96 dB
16 MHz to 3 GHz 106 dB
Reflection uncertainty
Magnitude
Transmission uncertainty
Magnitude
Measurement port characteristics
Frequency range
Corrected 30 kHz - 300 kHz
2
300 kHz - 1.3 GHz 1.3 GHz - 3 GHz
Directivity 48 dB 48 dB 43 dB
Source match 47 dB 41 dB 35 dB
Load match 48 dB 48 dB 43 dB
Reflection
±(0.004 dB ±(0.010 dB ±(0.019 dB
tracking
+0.02 dB/˚C) +0.01 dB/˚C) +0.02 dB/˚C)
Transmission
±0.018 dB ±(0.016 dB ±(0.033 dB
tracking
+0.02 dB/˚C) +0.01 dB/˚C) +0.02 dB/˚C)
Uncorrected
2
30 kHz - 300 kHz 300 kHz - 1.3 GHz 1.3 GHz - 3 GHz
Directivity 20 dB* 35 dB 30 dB
Source match 16 dB** 16 dB 16 dB
Load match 15 dB** 18 dB 16 dB
Reflection ±2.5 dB ±1 dB ±1 dB
tracking
Transmission ±2.5 dB ±1 dB ±1 dB
tracking
Crosstalk 90 dB*** 100 dB 100 dB
* 15 dB, 30 to 50 kHz **10 dB, 30 to 50 kHz *** 60 dB, 30 to 50 kHz
Phase
Phase
1. Typical below 300 kHz.
2. Typical performance.