SCALE OPERATION
CALIBRATION
Bar Code Calibration Procedure without Remote Display*
5. Reinstall the platter and power up the unit.
Figure 54. Platter Installation and Power Up
See power source caution statement on page 8 of this manual.
6. Enter bar code calibration mode. Use the vertical window to scan the following bar code. The scanner will beep once
as it enters the bar code calibration mode and the calibration utility will start.
³ 9 9 6 0 0 0
Figure 55. Entering Bar Code Calibration Mode
If a razz tone sounds, an error has occurred. Refer to Diagnostic Indicator Display; Error Codes starting on
page 30 for additional information.
7. Make sure there is no load on the scale platter. The white LEDs will be used to indicate the current step in the calibration
process.
8. Calibrate a zero load. The white LEDs will blink once periodically. Wait 8 to 10 seconds for scale stability then press the
beeper volume switch one time. The beeper will beep one time indicating that the Cal 1 value has been stored.
Figure 56.
* These procedures are for MS2320 StratosH Scanner/Avery Scale models only.
Bar Code Calibration Mode