44
SCALE OPERATION
CALIBRATION
Bar Code Calibration Procedure without Remote Display*
9. Place a half load on the center of the scale platter. The white LEDs will blink twice periodically. Wait for scale stability,
and then press the beeper volume switch once. The beeper will beep two times indicating that the Cal 2 value has been
stored.
Figure 57.
10. Add an additional half load to the existing half to simulate a full load (center the entire load). The white LEDs will blink
three times periodically. Wait for scale stability, and then press the beeper volume switch once. The beeper will beep
three times indicating that the Cal 3 value has been stored.
Figure 58.
11. Remove half of the load and center the remaining load. The white LEDs will blink four times periodically. Wait for scale
stability, and then press the beeper volume switch once. The beeper will beep four times indicating that the Cal 4 value
has been stored.
Figure 59.
12. Remove the remaining half load from the scale. The white LEDs will blink five times periodically. Wait for scale stability,
and then press the beeper volume switch once. The beeper will beep five times indicating that the Cal 5 value has been
stored.
Figure 60.
* These procedures are for MS2320 StratosH Scanner/Avery Scale models only.