Invacare SP Personal Computer User Manual


 
SECTION 10—BATTERIES
Part No 1143190 55 TDX™ SP
Replacing Batteries
NOTE:Forthisprocedure,refertoFIGURE 10.1onpage 56.
NOTE:Thefollowingtoolsarerequiredtoperformthisprocedure:
•WireCutter
1. Remove thebatteriesfromthewheelchair.RefertoRemoving/InstallingtheBatteries
From/IntotheWheelchaironpage 57.
2. Cutthetiewrapsthatsecurethebatteryterminalcoverstothebatteryterminals.
3. SlidetheREDbatteryterminalcoverbackontheREDbatterycabletoexposethe
POSITIVEbatteryterminal.
4. SlidetheBLACKbatteryterminalcoverbackontheBLACKbatterycabletoexpose
NEGATIVEbatteryterminal.
ƽ WARNING
NEVER allow any of your tools and/or battery cable(s) to contact BOTH battery
post(s) at the same time. An electrical short may occur and serious personal injury
or damage may occur.
5. Remove thelocknutthatsecuresthebracketofthePOSITIVEbatterycabletothe
POSITIVE(+)batterypostofthebattery.
6. Remove thelocknutthatsecurestheNEGATIVEbatterycabletotheNEGATIVE()
batterypostofthebattery
7. Discardtheexistingbattery.
8. Positionbatteryconnectorbracketorwiringharnessontothenew22NForGP24
batteryasshown.
9. SecuretheNEGATIVEbatterycabletotheNEGATIVE()batterypostwithexisting
mountingscrewandlocknut.
10. SecurethebracketofthePOSITIVEbatterycabletothePOSITIVE(+)batterypostwith
existingmountingscrewandlocknut.
11. Positioneachbatteryterminalcoverovertopofeachbatteryterminal.
12. Securebatteryterminalcoversinplacewithonetiewrap.
13. Installbatteriesintowheelchair.RefertoRemoving/InstallingtheBatteriesFrom/Into
theWheelchaironpage 57.