Metrologic Instruments MS2xxx Series Scanner User Manual


 
2
SCALE PROGRAM MODE
The following steps are only required for the scale configuration bar
codes located on pages 5 - 19 (bar codes beginning with 996.)
1. Power down the scanner/scale.
Then, temporarily remove the
scale platter and place it in a safe
location.
Figure 1.
2. Cut and remove the security seal over the program/calibration switch cover.
See Figure 2 below for seal location on the StratosE and the StratosH
models. See Figure 3 on page 3 for the seal location on the StratosS.
Once the security seal is cut the unit cannot be place in service until
it is calibrated, sealed and certified by local Weights and Measures
authority.
3. Remove the M3 screw securing the switch cover in place. Store the cover
and screw in a safe location. They will be needed again after the scale is
configured and calibrated.
Always follow all Electro-Static Discharge (ESD) procedures when
exposing the internal scanner/scale components.
4. Enter the scale program mode by sliding the switch to the program
position. See Figure 2 below for the switch location on the StratosE and
StratosH. See Figure 3 on page 3 for the switch location on the StratosS.
Figure 2. StratosE and StratosH Calibration Seal and Switch Location