National Instruments 6025E Switch User Manual


 
Chapter 4 Signal Connections
6023E/6024E/6025E User Manual 4-8 ni.com
Analog Input Signal Overview
The analog input signals for these devices are ACH<0..15>, ASENSE, and
AIGND. Connection of these analog input signals to your device depends
on the type of input signal source and the configuration of the analog input
channels you are using. This section provides an overview of the different
types of signal sources and analog input configuration modes. More
specific signal connection information is provided in the Analog Input
Signal Connections section.
Types of Signal Sources
When configuring the input channels and making signal connections,
you must first determine whether the signal sources are floating or
ground-referenced.
PFI4/GPCTR1_GATE DIO V
cc
+0.5 3.5 at (V
cc
-0.4) 5at0.4 1.5 50 kpu
GPCTR1_OUT DO 3.5 at (V
cc
-0.4) 5at0.4 1.5 50 kpu
PFI5/UPDATE* DIO V
cc
+0.5 3.5 at (V
cc
-0.4) 5at0.4 1.5 50 kpu
PFI6/WFTRIG DIO V
cc
+0.5 3.5 at (V
cc
-0.4) 5at0.4 1.5 50 kpu
PFI7/STARTSCAN DIO V
cc
+0.5 3.5 at (V
cc
-0.4) 5at0.4 1.5 50 kpu
PFI8/GPCTR0_SOURCE DIO V
cc
+0.5 3.5 at (V
cc
-0.4) 5at0.4 1.5 50 kpu
PFI9/GPCTR0_GATE DIO V
cc
+0.5 3.5 at (V
cc
-0.4) 5at0.4 1.5 50 kpu
GPCTR0_OUT DO 3.5 at (V
cc
-0.4) 5at0.4 1.5 50 kpu
FREQ_OUT DO 3.5 at (V
cc
-0.4) 5at0.4 1.5 50 kpu
AI = Analog Input DIO = Digital Input/Output pu = pullup
AO = Analog Output DO = Digital Output
Note: The tolerance on the 50 kpullup and pulldown resistors is very large. Actual value can range between 17 kand
100 k.
Table 4-3. I/O Signal Summary (Continued)
SignalName
Signal
Type and
Direction
Impedance
Input/
Output
Protection
(Volts)
On/Off
Source
(mA at V)
Sink
(mA
at V)
Rise
Time
(ns)
Bias